1-27 Model
The thermocouple model is made of the MI cable and an industrial head. It applies to measure temperature of gas-like and air environments. It's recommended to order the 4-20 mA / RS485 transducer (T-bus/ModBus protocol) in the remote Z67 enclosure with a cable gland and connect it by the compensating cable to avoid overheating.

It may be bent in any direction
NOTE! Temperature near the industrial head must stay up to 300°С. The industrial head ensures IP54 protection only up to 120°С.
Technical specifications
| Type and Nominal Resistance | Type К Type N |
| Accuracy | 1, 2 |
| Junction type | Ungrounded |
| Industrial head | A (by default) B (on order) |
| MI cable | Stainless steel AISI 321 Ø3 mm, stainless steel AISI 321 Ø6 mm, stainless steel 310 Ø6 mm, nickel alloy Inconel 600 (INC) Ø4.5 mm, nickel alloy Inconel 600 (INC) Ø6 mm, nickel-chromium alloy Nicrobel (NIC) Ø6 mm |
| Process temperature range, °С | 0...800 (type K AISI 321) 0...1050 (type K steel 310) 0...1100 (type K INC) 0...1150 (type N INC) 0...1200 (type N NIC) |
| Thermal inertia, sec | 45 |
Dimensions
| Length of the immersion part, L, mm | 250, 320, 400, 500, 630, 800, 1000, 1250, 1500, 2000... up to 20 000 |
| Diameter of the immersion part, mm | Ø3, Ø4.5, Ø6 |